Dijamin original dan bergaransi
TSN 807 IWSC
Power supply AC 220 V, 50 Hz, 700 Watt
Dimension Appr. 71 L x 110 W x 185 H cm
Weight Appr. 120 kg
Castors 4 inches, two lockable
Mattress tray ABS
Mattress size Appr. 64 cm x 41 cm
Mattress elevation 0 to 12° by gas spring
Mattress height from ground 100-120 cm (for type with height adjusment)
Treatment distance Appr. 75 cm
IV Pole Adjustable with fixation clamp
Tray For instrument with max.load 2.5 kg
Head part Rotatable
X-Ray cassette tray Acrylic
Untuk informasi harga dan yang lainnya, silakan hubungi kami
TSN 807 IWSC
Power supply AC 220 V, 50 Hz, 700 Watt
Dimension Appr. 71 L x 110 W x 185 H cm
Weight Appr. 120 kg
Castors 4 inches, two lockable
Mattress tray ABS
Mattress size Appr. 64 cm x 41 cm
Mattress elevation 0 to 12° by gas spring
Mattress height from ground 100-120 cm (for type with height adjusment)
Treatment distance Appr. 75 cm
IV Pole Adjustable with fixation clamp
Tray For instrument with max.load 2.5 kg
Head part Rotatable
X-Ray cassette tray Acrylic
Control panel type Hidden keys type, auto self-test system, LED seven segment
Control mode Servo (Skin mode) and manual (Air mode)
Controller type Microcontroller
Control panel lock Automatic key lock after 10 seconds unchanged
Temperature display range 17.1°C - 42.4°C
Air mode control range 25.0°C - 37.0°C, Override 39.0°C, Resolution 0.1°C
Skin mode control range 34.0°C - 37.0°C, Override 38.0°C, Resolution 0.1°C
Radiant heat source Quartz Infrared heater 600 Watt
Warm-up time ± 30 minutes
Illumination 2 x 10 W fluorescent lamps
Alarm indicator Visible (dot-matrix LCD) and audible (buzzer)
Alarm silencer 2 minutes for low internal battery and mains failure.
10 minutes for oyhers.
Information on LCD Heater level, alarms, serial number, internal battery level
and User ID (by customer request)
• System failure
• High temperature
• Low internal battery
• Skin over temperature
• Sensor failure
• Air over temperature
• Mains failure
• Warm-up failure (>30 minutes)
• Probe failure (probe disconnected)
• Skin deviation ± 1°C and air deviation ± 3°C
• APGAR timer (1 st, 5th and 10th minute)
• Adjustable height • O2 tank 1000 liters capacity + regulator • Oxyhood • Phototherapy lamp (1x20 W blue light lamp)