Dijamin original dan bergaransi
Accept syringe from 2 to 60 ml
The Aitecs 2016 syringe pump is designed to meet the fluid and drug delivery requirements of today’s changing clinical environment including general wards, critical and intensive care, neonatal, operating rooms and emergency rooms. The Aitecs 2016 offers enhanced safety, reliability, ease of use, and flexibility.
• Accept syringe from 2 to 60 ml
• Customized drug library ( up to 350 drug labels)
• Highly visible pump status indicator
• Easy intutive programing and operation
• Improved alarm signal
• Graphic LCD DIsplay
• Rate range from 0.01 to 2200 ml/h
• Advanced communication ( USB , IrDA )
• Line disconnection alarm.
• Rapid occulsion detection ROD
• Can be mounted on : Universal pole clamp < infusion device rack
• Dose Unit : g, mg, mcg, ng, Units, Kunits, mUnits , mcUnits, mol, mmol , mcmol,
mEq, and other configured by the medical personnel.
• Dose error reduction : infugard Protocal ( with safety limits )
Specification
Flow rate : 0.01 - 99.99 ml/h in steps of 0.01 ml/h
100.0 - 999.9 ml/h in steps of 0.1 ml/h
1000 - 2200 ml/h in steps of 1 ml/h
Mechanical accuracy : ± 1%
Volmetric accuracy approved syringes : ± 2% ( for 1 ml/h above , over long periods by
EN 60601-2-24
Compatible syringes , ml : All mains brands in size 2/3 5/10,20,30,50/60
Occulusion pressure ( adjustable)
: L1 ( app 50 mmHg )
L3 ( app 250 mmHg)
L5 ( app. 450 mmHg)
L7 ( app. 650 mmHg)
L10 ( app. 850 mmHg )
Purge volume
: 0,1 - 4.0 ml in 0.1 ml steps
Bolus volume
: 0,1 - 60.0 ml in 0.1 steps
Bolus / purge rate
:
1-2200 ml/h in 0.1 ml/h steps
Keep Vein Open (KVO) rate
:
For syringes 5 ml and below 0.01 - 5.00 ml/h
For syringe 10 ml and above 0.10 -5.00 ml/h
KVO volume
: 0.1 % - 10% of the syringe volume
Volume to be infused (VTBI)
: 0.1 - 99.9 ml in steps of 0.01 ml
100 - 999 ml in steps of 0.1 ml
1000 - 9999 ml in steps of 1 ml or no Limit
Alarms : Visual and audio
Selectable options for customized
: Various
operation
Power supply
:
100-240 VAC, 50/60 Hz
or Internal rechargeable battery
Post VTBI rate
: Stop KVO, or continue with set rate
Battery type
: Built-In NiMH 7.2 V/ 2.5 Ah
Battery life
: 10 hours @ 5 ml/h
Battery charging time
: 5 hours to 100% charge
Protection : Class II , IPX1
Standby time
:
1 min - 24 h
Even log
:
> 2000 events
History Log
: > 500 events
Weight
: 2.3 Kg
Dimension :
120 x 320 x 137 mm
Key log
: > Last 300 press