Dijamin original dan bergaransi
The LogTag® TRID30-7 Temperature Recorder measures and stores up to 7770 temperature readings over a measurement range of -30°C to +60°C (-22°F to +140°F).
The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.
Models | TRID30-7R : (replaceable battery, download instructions here) TRID30-7F : (fixed battery) TRID30-7FW (fixed battery, preconfigured and package to meet WHO specification E006/TR06.3) |
Measurement Range | -30°C ~ +60°C (-22°F ~ +140°F) |
Resolution | 0.1°C(0.2°F) for measurements -30°C ~ +40°C (-22°F ~ +104°F) 0.2°C(0.4°F) for measurements above +40°C (+104°F) |
Accuracy | Better than ±0.5°C (±0.9°F) for measurements from -20°C~+40°C (-4°F~+104°F) -typically ±0.3°C (0.6°F) Better than ±0.8°C (±1.5°F) for other measurements - typically ±0.5°C (0.9°F). |
Sensor Reaction Time | Typically less than 5 minutes (T90) in moving air (1m/s). |
Capacity | Data logging memory : 7770 logs ( 53 days @ 10min logging, 80 days @ 15min logging) Day summary statistics memory (for display on LCD): up for 30 days of Max/Min and duration values |
Sampling Frequency | Adjustable, 30 seconds to several hours |
Logging Start Options | Push button start or specific date & time. Optional start delay of up to 18hours. |
Recording Indication | State indicator "RECORDING" |
Download Time | Typically with full memory (7770 readings) in less than 5 seconds depending on computer or readout device used. |
Environmental | IP65 (roughly equivalent to NEMA 4) when vertically mounted or hung. |
Power Source | 3V Lithium-Manganese Dioxide extended temperature chemistry |
Battery Life | TRID30-7R: typically 1 year of operation TRID30-7F: typically 2+ years of operation NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily for no longer than 30 seconds each time. Download data monthly. |
Size | 93mm(H)x54.5mm(W)x8.6mm(T) |
Weight | TRID30-7R: 39grams TRID30-7F: 43grams |
Case Material | Polycarbonate |
Other Features | • Optional clearing of alert indication by push button (places inspection make at same time). • Download inspection mark recorded in LogTag memory every time LogTag is downloaded. • 'Prestart' fail-safe logging (records temperature data even if not started) • Low Battery indication in software. • Comprehensive Alarm trigger configuration. • Calibration to achieve higher accuracy possible. |